Customer Support: 131 242

  • There are no items in your cart
We noticed you’re not on the correct regional site. Switch to our AMERICAS site for the best experience.
Dismiss alert

ASTM D 7363 : 2013

Superseded
Superseded

A superseded Standard is one, which is fully replaced by another Standard, which is a new edition of the same Standard.

View Superseded by
superseded

A superseded Standard is one, which is fully replaced by another Standard, which is a new edition of the same Standard.

Standard Test Method for Determination of Parent and Alkyl Polycyclic Aromatics in Sediment Pore Water Using Solid-Phase Microextraction and Gas Chromatography/Mass Spectrometry in Selected Ion Monitoring Mode
Available format(s)

Hardcopy , PDF

Superseded date

11-11-2014

Language(s)

English

Published date

15-03-2013

CONTAINED IN VOL. 11.02, 2018 Defines the separation of pore water from PAH-impacted sediment samples, the removal of colloids, and the subsequent measurement of dissolved concentrations of the required 10 parent PAHs and 14 groups of alkylated daughter PAHs in the pore water samples.

1.1The U.S. Environmental Protection Agency (USEPA) narcosis model for benthic organisms in sediments contaminated with polycyclic aromatic hydrocarbons (PAHs) is based on the concentrations of dissolved PAHs in the interstitial water or “pore water” in sediment. This test method covers the separation of pore water from PAH-impacted sediment samples, the removal of colloids, and the subsequent measurement of dissolved concentrations of the required 10 parent PAHs and 14 groups of alkylated daughter PAHs in the pore water samples. The “24 PAHs” are determined using solid-phase microextraction (SPME) followed by Gas Chromatography/Mass Spectrometry (GC/MS) analysis in selected ion monitoring (SIM) mode. Isotopically labeled analogs of the target compounds are introduced prior to the extraction, and are used as quantification references.

1.2Lower molecular weight PAHs are more water soluble than higher molecular weight PAHs. Therefore, USEPA-regulated PAH concentrations in pore water samples vary widely due to differing saturation water solubilities that range from 0.2 µg/L for indeno[1,2,3-cd]pyrene to 31 000 µg/L for naphthalene. This method can accommodate the measurement of microgram per litre concentrations for low molecular weight PAHs and nanogram per litre concentrations for high molecular weight PAHs.

1.3The USEPA narcosis model predicts toxicity to benthic organisms if the sum of the toxic units (ΣTUc) calculated for all “34 PAHs” measured in a pore water sample is greater than or equal to 1. For this reason, the performance limit required for the individual PAH measurements was defined as the concentration of an individual PAH that would yield 1/34of a toxic unit (TU). However, the focus of this method is the 10 parent PAHs and 14 groups of alkylated PAHs (Table 1) that contribute 95 % of the toxic units based on the analysis of 120 background and impacted sediment pore water samples.3 The primary reasons for eliminating the rest of the 5-6 ring parent PAHs are: (1) these PAHs contribute insignificantly to the pore water TU, and (2) these PAHs exhibit extremely low saturation solubilities that will make the detection of these compounds difficult in pore water. This method can achieve the required detection limits, which range from approximately 0.01 µg/L, for high molecular weight PAHs, to approximately 3 µg/L for low molecular weight PAHs.

TABLE 1 Target PAHs, Toxic Unit Factors and Performance LimitsA

Analyte

Added d-PAH
Internal
Standard

d-PAH Internal
Std. for Calculation

Conc. for One
Toxic Unit,
Ctu, (ng/mL)

Performance Limit
(ng/mL)

Basis for
Performance
LimitB

Naphthalene

A

A

193.47

5.69

B

2-Methylnaphthalene

B

81.69

2.40

B

1-Methylnaphthalene

B

B

81.69

2.40

B

C2-Naphthalenes

A

30.24

0.89

B

C3-Naphthalenes

A

11.10

0.33

B

C4-Naphthalenes

A

4.05

0.12

C

Acenaphthylene

C

308.85

9.03

B

Acenaphthene

C

C

55.85

1.64

B

Fluorene

D

D

39.30

1.16

B

C1-Fluorenes

D

13.99

0.41

B

C2-Fluorenes

D

5.30

0.16

B

C3-Fluorenes

D

1.92

0.06

S

Phenanthrene

E

E

19.13

0.56

B

Anthracene

E

20.72

0.61

B

C1-Phenanthrenes/Anthracenes

E

7.44

0.22

B

C2-Phenanthrenes/Anthracenes

E

3.20

0.09

B

C3-Phenanthrenes/Anthracenes

E

1.26

0.04

B

C4-Phenanthrenes/Anthracenes

E

0.56

0.02

S

Fluoranthene

F

7.11

0.21

B

Pyrene

F

F

10.11

0.30

B

C1-Fluoranthenes/Pyrenes

F

4.89

0.14

C

Benz[a]anthracene

G

2.23

0.066

B

Chrysene

G

G

2.04

0.060

B

C1-Chrysenes/Benz[a]anthracenes

G

0.86

0.025

C

A From Hawthorne, S. B., Grabanski, C. B., Miller, D. J., and Kreitinger, J. P., “Solid Phase Microextraction Measurement of Parent and Alkyl Polycyclic Aromatic Hydrocarbons in Milliliter Sediment Pore Water Samples and Determination of KDOC Values,” Environmental Science Technology, Vol 39, 2005, pp. 2795–2803.
B Performance limits were determined as 3 times the background concentrations from the SPME fiber based on the analysis of water blanks (“B”), the lowest calibration standard which consistently yielded a signal to noise ratio of at least 3:1 (“C”), or (for when no calibration standard was available) for the lowest concentrations consistently found in pore water samples with a signal to noise ratio of at least 3:1 (“S”). Detection limits for alkyl PAHs are based on a single isomer.

1.4The test method may also be applied to the determination of additional PAH compounds (for example, 5- and 6-ring PAHs as described in Hawthorne et al.).4 However, it is the responsibility of the user of this standard to establish the validity of the test method for the determination of PAHs other than those referenced in 1.1 and Table 1.

1.5This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. For specific hazard statements, refer to Section 9.

Committee
D 19
DocumentType
Test Method
Pages
24
PublisherName
American Society for Testing and Materials
Status
Superseded
SupersededBy
Supersedes

ASTM E 3163 : 2018 Standard Guide for Selection and Application of Analytical Methods and Procedures Used during Sediment Corrective Action
ASTM E 3164 : 2018 Standard Guide for Sediment Corrective Action – Monitoring

ASTM D 2777 : 2013 : REDLINE Standard Practice for Determination of Precision and Bias of Applicable Test Methods of Committee D19 on Water
ASTM D 1192 : 1998 Standard Guide for Equipment for Sampling Water and Steam in Closed Conduits (Withdrawn 2003)

View more information
$176.78
Including GST where applicable

Access your standards online with a subscription

Features

  • Simple online access to standards, technical information and regulations.

  • Critical updates of standards and customisable alerts and notifications.

  • Multi-user online standards collection: secure, flexible and cost effective.

Need help?
Call us on 131 242, then click here to start a Screen Sharing session
so we can help right away! Learn more